There are four types of gold yellow, white, red (pink) and green.
Diagram 1 -Display Unit
- Electronic unit
- Alpha-numeric Display Window
- Low Battery Indicators
- Gold Selection Buttons
- Sensor Outlet
- Optional Converter (Connection)
- ON-OFF Power Switch
Note: When using the optional converter, the power switch must be in the "OFF" position.
The Mechanical Sensor
The Mechanical Sensor for the G-XL-18 Gold Tester consists of the sensor body and a polyethylene tube with gel. The polyethylene tube must be placed inside of the sensor's body before beginning a test.
Diagram 2 shows each part of the mechanical sensor.
Diagram 2 - Sensor
When using the Sensor, be sure that:
- The sensor nozzle is clean. Use the plastic cleaner (supplied) to clean the opening.
- When testing, dispense THEN dispose of the drop of gel following each test. For best result, use every other gel drop for testing.
Diagram 3 - Polyethylene Tube
- Tube Cap
- Tube Nozzle
- Tube Body
- Feed Screw
- Piston Engagement
- Bottom Plug
- Feed Screw
Loading/Changing the Polythylene Gel Tube
- Take off the nozzle cap (Diagram 2).
- Remove the Stationary Cap by turning the light colored barrel (Diagram 2) counter clockwise. Please remember that the Stationary Cap and the Rotary Cap are undetachable. (Diagram 2).
- Insert the Feed Screw Engagement (Diagram 3, #7) into the appropriate slot of the Rotary Cap. Be sure that the Feed Screw is fully engaged.
- Take off tube Cap (Diagram 3, #1)
- Insert the polyethylene gel tube, which is fully lodged in the Stationary/Rotary Cap, into the sensor and turn the bright colored barrel (Diagram 2) clockwise until fully seated. Do not turn the dark colored top barrel (Diagram 2, # 7) clockwise or it will damage the mechanism.
- Turn the Rotary Cap counter clockwise several times until the first drop of gel appears at the sensor nozzle.
Operation Procedures
- Load the polyethylene gel tube.
- Connect the black wire to the sensor's plug outlet, the black and red wire connector to the display unit and the red wire to the testing plate.
- Turn the unit "ON". (See Diagram 1, #7 for switch). The unit should display: "G-XL-18 READY TO TEST…"
- Select a testing point on the object close to, BUT NOT TOUCHING, the alligator clip.
- Holding the sensor in the vertical position with the nozzle down, twist the Rotary
Cap counter clockwise, a click at a time, until a drop of gel appears on the tip. Wipe
this first drop of gel on the towel and turn (usually one or two clicks) until a small
gel sample appears. Immediately touch the sensor to pre-selected cleaned point on
the test object.
DO NOT LET GEL OR SENSOR NOZZLE TOUCH THE ALLIGATOR CLIP! - Press one of the following buttons: Y-Yellow, W-White, R-Red, or G-Green depending on the color of the gold (See Diagram 1, #4 for button location).
- Maintain the sensor in the vertical position for approximately 5 seconds while the instruments calibrates the gold value.